Bridge power supply
|
DC5 V±0.25 V within 60 mA
(changeable to DC2.5 V)
|
Input range
|
F.S. setting is available with the input range from 0.4 mV/V to 3.1 mV/V. (When the bridge power supply is DC5 V)
|
Output range
|
DC5 V Load resistance is 5 kΩor more.
(F.S. setting is available by function)
|
Output times
|
Changeable to 4 times/s or 20 times/s
(Synchronous with display rate)
|
Output resolution
|
Approx. 1/10 000
|
Zero adjustment range
|
-0.3 mV/V ~ 2.4 mV/V
|
Non-linearity
|
Display 0.05 %F.S., Output 0.05 %F.S.
|
Temperature efficient
|
Zero point ±1 μV/℃, Sensitivity ±0.02 %F.S./℃
(Input conversion, During F.S. setting at the input from 0.5 mV/V to 3.1 mV/V)
|
Input noice
|
±1 μVp-p or less
(At the default setting of the digital filter and the stabilized filter)
|
Input filter
|
1 Hz (At the "0" setting of the digital filter and the stabilized filter)
|
A/D sampling
|
20 times/s
|
Load display
|
|
Display range
|
-9 999 ~ 99 999
|
Display times
|
4 times/s (Changeable to 20 times/s)
|
Decimal point display
|
Changeable to No display, 101, 102, 103, 104
|
External control function
|
ZERO, HOLD, PEAK/TRACK/A/Z, RESET/A/Z OFF
|
Comparator function
|
|
Set value
|
-99 999 ~ 99 999
|
Set quantity
|
2 points of S1 and S2
|
Comparator conversion rate
|
Changeable to 4 times/s or 20 times/s
(Synchronous with display rate)
|
Contact output function
|
2 points of S1 and S2
|
Operating temp./humidity range
|
Temperature 0 ℃ ~ 50 ℃, Humidity 85 %RH or less
(Non condensing)
|
Power supply voltage
|
AC100 V ~ AC240 V 50/60 Hz
(Permissible variable range AC85 V ~ AC264 V)
|
Power consumption
|
Approx. 11 VA (without options, at AC100 V)
Approx.14 VA at maximum (with options, at AC100 V ~ 240 V)
|
Outline dimensions(WxHxD)
|
96 mm x 48 mm x 120 mm (excludes protruding parts)
|
Weight
|
Approx. 0.3 kg (without options)
|
Protection code
|
IP65 (Only for the fron panel when the attached panel mounting gascket is installed.
|