Bridge power supply
|
DC10 V±0.5 V within 60 mA
(Changeable to DC5 V or DC2.5 V)
|
Input range
|
F.S. setting is available at the input range from ±0.4 mV/V to ±3.1 mV/V (When bridge power supply is DC10 V or DC5 V)
F.S. setting is available at the input range from ±0.8 mV/V to ±3.1 mV/V (When bridge power supply is DC2.5 V)
|
Output range
|
DC±10 V Load resistance 5 kΩor more
(F.S. setting is available in the function.)
|
Output times
|
100 times/s, 500 times/s, 1 000 times/s, 2 000 times/s (Synchronous with A/D sampling rate.)
|
Output resolution
|
Same as the display resolution.
|
Zero adjustment range
|
±2.0 mV/V
|
Non-linearity
|
Display 0.025 %F.S.
Output 0.025 %F.S.
|
Temperature coefficient
|
Zero point ±0.5 μV/℃ (Input conversion, in F.S. setting at the input of ±0.5 mV/V ~ 3.1 mV/V)
Sensitivity ±0.01 %F.S./℃ (Input conversion, F.S. setting at the input of ±0.5 mV/V ~ 3.1 mV/V)
|
Input filter
|
10 Hz, 100 Hz, 1 kHz
|
A/D sampling
|
2 000 times/s
(Changeable to 100 times/s, 500 times/s, 1 000 times/s)
|
Load display
|
-11 000 ~ 11 000
|
Comparator display
|
-99 999 ~ 99 999
|
Nos. of display
|
1 point (Selectable from S0, S1, S2, S3 or S4)
|
Display times
|
20 times/s (Changeable to 4 times/s, 50 times/s or 100 times/s)
|
External control function
|
ZERO, PEAK/TRACK, HOLD, RESET, SEL.1, SEL.2
|
Comparator function
|
|
Display range
|
-99 999 ~ 99 999
|
Nos. of display
|
1 point (Selectable from S0, S1, S2, S3 or S4)
|
Display times
|
20 times/s (Changeable to 4 times/s, 50 times/s or 100 times/s)
|
Open collector output function
|
5 points of S0, S1, S2, S3 or S4
|
Operating temp./humidity range
|
Temperature -10 ℃ ~ 50 ℃, Humidity 85 %RH or less
(Non condensing)
|
Power supply voltage
|
AC100 V ~ AC240 V 50/60 Hz
(Permissible variable range AC85 V ~ AC264 V)
|
Power consumption
|
Approx.15 VA (without options, at AC100 V)
Approx.18 VA at maximum (with options, at AC100 V ~ 240 V)
|
Outline dimensions(WxHxD)
|
96 mm x 96 mm x 129.5 mm (excludes protruding parts)
|
Weight
|
Approx. 500 g (without options)
|
Protection code
|
IP65 (Only for the fron panel when the attached panel mounting gascket is installed.
|