Model
|
TGI-250kN
|
TGI-100kN
|
TGI-50kN
|
TGI-10kN
|
TGI-5kN
|
TGI-1kN
|
TGI-500N
|
TGI-200N
|
Maximum test force
|
250 kN
|
100 kN
|
50 kN
|
10 kN
|
5 kN
|
1 kN
|
500 N
|
200 N
|
Test force measure -ment range *1)
|
0.1N~
250kN
|
0.1N~
100kN
|
0.1N~
50kN
|
0.1N~
10kN
|
0.1N~
5kN
|
0.1N~
1kN
|
0.1N~
500N
|
0.1N~
200N
|
Test space
|
Tensile / compression testing beneath the movable crosshead.
(A model is available allowing tensile testing above the movable crosshead (Optional).)
|
Loading method
|
Constant-speed strain loading with highly accurate AC servo motor and backlash-free precise ball screw.
|
Cross head speed / permissible test force
|
Within the max. test force over the entire speed ranges.
|
Test force range
|
2 methods:
AUTO: (accuracy guranteed for 1/1 to 1/250 of load cell capacity)
100: (accuracy guranteed for 1/100 to 1/1 000 of load cell capacity)
|
Test force display
|
5-digits digital display ±99,999
AUTO range: Resolution 1/5000, test force reduction increase/decrease, Full scale display, and automatic switching of decimal point position and display unit.
100 range: Resolution 1/10000
* Test force display varies depending on the load cell provided.
|
Test force measuring accuracy (Guaranteed accuracy range)
|
Standard accuracy : ±1 % of indicated value
(1/1 to 1/1000 of load cell capacity)
High accuracy : ±0.5% of indicated valeu (Optional)
(1/1 to 1/1000 of load cell capacity)
* Test machine must be in location free of vibration
|
Applicable standard
|
JIS B 7721 Class 1 or Class 0.5 and ISO7500-1 Class 1 or Class 0.5
|
Effective test width (Note 2) (mm)
|
600
|
600
|
420
|
430
|
Cross head stroke (mm)
|
1 550
|
1 350
|
1 015
|
1 090
|
Maximum mounting space (mm)
|
1 690
|
1 555
|
1 165
|
1 160
|
Tensile stroke (Note 3) (mm)
|
620
|
850
|
470
|
710
|
820
|
870
|
890
|
910
|
Crosshead speed
|
0.000 5 to 500 mm/min
0.1 to 500 mm/mincan be set in 0.1 mm/min incliments. Below this, can be set in 9 steps: 0.05, 0.03, 0.02, 0.01, 0.005, 0.003, 0.002, 0.001, and 0.0005.
|
0.000 5 to 1 000 mm/min
0.1 to 1 000 mm/min; can be set in
0.1 mm/min increments. Below this, can be set in 9 steps: 0.05, 0.03, 0.02, 0.01, 0.005, 0.003, 0.002, 0.001, and 0.0005.
|
Crosshead speed accuracy
|
±0.1 % over entire speed range
|
Return speed
|
Can be set as required between 0.1 mm/min and 750 mm/min
|
Can be set as required between 0.1 mm/min and 1 500 mm/min
|
Speed setting method
|
Manual: Numeric input via control panel
When using data processing software: Automatically set in accordance with test conditions created.
|
Manual speed
|
3 speeds (HIGH/MIDDLE/LOW); each freely settable
|
Test mode
|
Single and cycle (Combination of test force and position)
Optional: Creep, Automatic loading
|
Test force display
|
5 digit digital display ±99 999
Auto range: Resolution 1/5 000, test force reduction increase/decrease, full scale display, and automatic switching of decimal point position and display units
100 range: Resolution 1/10 000 * Test force display varies depending on the load cell provided.
|
Movable crosshead position display
|
6 1/2-digit digital display ±1999.999
|
Standard function
|
Load cell model & capacity
|
:
|
Auto adjust
|
Load cell span
|
:
|
Auto set for tension & compression
|
Zero point
|
:
|
One push
|
Position reset
|
:
|
One push
|
Sample rupture detection
|
:
|
Selectable from stop, reverse and ignore
|
Rupture sensitivity
|
:
|
Set as desired to 0.04 % to 99.00 % of test force
|
Test condition registration function
|
:
|
10 conditions
|
Test force & position analog output
|
:
|
DC0 ~ DC5 V (with scaling function)
|
Interface
|
:
|
Conforms to USB interface 2.0/1.1 standard (B- type connector)
|
|
Options
|
Chuck adapter
|
:
|
Lower chuck mounting adapter for tensile testing
|
Load cell
|
:
|
for tension / compression type
(with universal joint, connect pin and compression plate
----------
for tensile type
(with universal joint and connecting pin)
----------
for compression type
(with for compression plate)
|
Testing jig
|
:
|
Former jig is attachable.
|
Displacement measuring function
|
:
|
Strain input type
|
Test machine stand
|
Data processing software
|
:
|
Standard (Tensile / compression / bending), Peeling, Cycle (One way, Both way),
Stress relaxation
|
|
External dimensions (WxHxD)(mm)
|
1420 x 2406 x 894
|
1221 x 2192 x 720
|
900 x 1507 x 490
|
700 x 1583 x 523
|
700 x 1 583 x 513
(without level adjuster)
|
Approximate weight (kg)
|
1 200
|
900
|
310
|
120
|
Power supply voltage, Frequency
|
3-phase AC200 V to 230 V+10 % to 15 %, 50/60 Hz
|
Single phase AC100 V to 115 V+10% to -15%, 50/60 Hz
|
Power consumption (W)
|
7.5 k
|
2.9 k
|
1.8 k
|
750
|
400
|
Operating temperature / humidity range
|
Temperature 5 ℃ ~ 40 ℃,
Humidity 20 % ~ 80 %RH (with no condensation)
|